The REKKER FOX-R20 SSL (Former name REKKER DMH-002) gamma beam irradiator (GBI) system is to be utilized mainly for calibration of detectors and other radiation protection unit. The irradiator is a Category II. irradiator. The sealed sources are kept in a dry storage position (in fully shielded condition) when there are not in use (off position), the GBI shall be shielded on all sides.
Shielding
REKKER FOX-R20 SSL gamma beam irradiator (GBI) is made up from a lead shield, a welded-frame, and the additional equipment. All of them are under the cover. Same parts of the torpedo is made up from tungsten to minimalize the radiation radiation under the equipment.
Safety Features
The irradiator system is designed to be a safe system. “Fail Safe” design constraints have been applied to all components that involve source exposure. A safety interlock system is an integral part of the design to prevent inadvertent exposure.
In the case of signal from interlock the irradiator is stopped automatically and source goes back to its safe storage position. The interlocks work independently from each other and do not affect each other (diversity). In case of an aggressive access the detectors get the controller to break the irradiation immediately.
Radioactive Source
The FOX-R20 SSL gamma beam irradiator (GBI) is loaded with one double capsulated Cs-137 source, their activity maximum 7,4 TBq (20 Ci). The radioactive Cs-137 sources are doubly encapsulated – hermetically sealed in a stainless-steel capsule – special form sources.
Collimator
The collimator port of REKKER FOX-R20 SSL gamma beam irradiator (GBI) holds a removable collimator that is designed based on ISO 4037:1996 to provide a circular radiation beam. The Beam centerline is selectable by the customer but based on the facility (room) design we could offer the best solution.
Attenuator
A REKKER-M-1000 manual or REKKER-A-1000 automatic attenuator is installed in the system. Include set of primary beam attenuators (number of layers is 4-8 ) allowing to decreas the beam intensity by factor in 2-1000 range, installed on a rigid support frame. The outer attenuators contains a last tungsten shape for better beam profile. The attenuator level is adjusted manually or automatically by the operator before the irradiation
Control system (safety and security)
The irradiator operates through a remote-control system that is installed in the control room and has connection and indication to various interlocks and the components. It shows and collects every information from the subsystem and storages it. More buttons (start/stop/reset/emergency) are installed in the control panel. In the case of signal from interlock the irradiator is stopped automatically and source goes back to its safe storage position.